目的:研究影响机械通气患者BIS值的相关因素。方法:42例机械通气患者,BIS模块连续监测镇静深度48小时,记录患者镇静前后、每隔16小时的呼吸循环指标,BIS值、SAS评分,建立混合线形统计模型,评价BIS监测的影响因素。结果:患者的血压、心率、呼吸频率、PaO2、SPO2 、FiO2等均对BIS值无影响,仅有PaCO2和SAS评分与BIS相关(P<0.05)。结论:BIS模块监测机械通气患者镇静深度,BIS值的变化与患者PaCO2及SAS评分有关。
Citation:
LI Xiaojin,ZHANG Chuan,ZHENG Bixia,et al. The Research of Related Factors in Affecting the Bispectral Index Monitoring for the Depth of Sedation of Ventilated Patients. West China Medical Journal, 2009, 24(2): 362-363. doi:
Copy
Copyright © the editorial department of West China Medical Journal of West China Medical Publisher. All rights reserved
1. |
Brocas,E,Dupont,H,Paugam Burtz C.Bispectral index variations during tracheal suction in mechanically ventilated critically ill patients:effect of an alfentanil bolus[J].Intensive Care Med,2002,28(2):211-213..
|
2. |
Ball J.How useful is the bispectral index in the management of ICU patients?[J]Minerva Anestesiol,2002,68(4):248-251..
|
3. |
Simmons L E,Riker R R,Prato B S,et al.Assessing sedation during intensive care unit mechanical ventilation with the Bispectral Index and the SedationAgitation Scale[J].Crit Care Med,1999,27(8):1499-1504..
|
4. |
Mondello E,Siliotti R,Noto G,et al.Bispectral Index in ICU:correlation with Ramsay Score on assessment of sedation level[J].J Clin Monit Comput,2002,17(5):271-277..
|
5. |
Riker R R,Fraser G L,Simmons L E,et al.Validating the SedationAgitation Scale with the Bispectral Index and Visual Analog Scale in adult ICU patients after cardiac surgery[J].Intensive Care Med,2001,27(5):853-858..
|
6. |
张川,李孝锦,李晓玉,等.BIS监测对诊断机械通气患者镇静深度的价值[J].华西医学,2008,23,(1):41-42..
|
7. |
Medicis JJ,Hu A,Lopez C,et al.Comparsion of standard sedation assessment to bispectral analysis in patients who are on neuromuscular blockers:A pilot study[C].ASHP Midyear Clinical Meeting(ASHP Midyear Clinical Meeting),2002,37(DEC):P-347R..
|
- 1. Brocas,E,Dupont,H,Paugam Burtz C.Bispectral index variations during tracheal suction in mechanically ventilated critically ill patients:effect of an alfentanil bolus[J].Intensive Care Med,2002,28(2):211-213..
- 2. Ball J.How useful is the bispectral index in the management of ICU patients?[J]Minerva Anestesiol,2002,68(4):248-251..
- 3. Simmons L E,Riker R R,Prato B S,et al.Assessing sedation during intensive care unit mechanical ventilation with the Bispectral Index and the SedationAgitation Scale[J].Crit Care Med,1999,27(8):1499-1504..
- 4. Mondello E,Siliotti R,Noto G,et al.Bispectral Index in ICU:correlation with Ramsay Score on assessment of sedation level[J].J Clin Monit Comput,2002,17(5):271-277..
- 5. Riker R R,Fraser G L,Simmons L E,et al.Validating the SedationAgitation Scale with the Bispectral Index and Visual Analog Scale in adult ICU patients after cardiac surgery[J].Intensive Care Med,2001,27(5):853-858..
- 6. 张川,李孝锦,李晓玉,等.BIS监测对诊断机械通气患者镇静深度的价值[J].华西医学,2008,23,(1):41-42..
- 7. Medicis JJ,Hu A,Lopez C,et al.Comparsion of standard sedation assessment to bispectral analysis in patients who are on neuromuscular blockers:A pilot study[C].ASHP Midyear Clinical Meeting(ASHP Midyear Clinical Meeting),2002,37(DEC):P-347R..